Rosemount 848T Spécifications Page 12

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 18
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 11
December 2014
12
Quick Start Guide
Product Certifications
Approved Manufacturing Locations
Rosemount Inc. – Chanhassen, Minnesota, USA
Emerson Process Management GmbH & Co. - Germany
Emerson Process Management Asia Pacific Private Limited - Singapore
Telecommunication Compliance
All wireless devices require certification to ensure that they adhere to regulations
regarding the use of the RF spectrum. Nearly every country requires this type of
product certification. Emerson is working with governmental agencies around
the world to supply fully compliant products and remove the risk of violating
country directives or laws governing wireless device usage.
FCC and IC
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following conditions: This device may not cause harmful interference. This device
must accept any interference received, including interference that may cause
undesired operation.
This device must be installed to ensure a minimum antenna separation distance
of 20 cm from all persons.
Ordinary Location Certification for FM
As standard, the transmitter has been examined and tested to determine that the
design meets basic electrical, mechanical, and fire protection requirements by
FM, a nationally recognized testing laboratory (NRTL), as accredited by the
Federal Occupational Safety and Health Administration (OSHA).
Hazardous Locations Certificates
North America
Factory Mutual (FM) approvals
N5 FM Class 1, Division 2, and Dust
Certificate: 3034378
Applicable Standards: Class 3600:1998, Class 3610:2010, Class 3810:2005,NEMA 250:
1997,ANSI/ISA-60079-0:2009, ANSI/ISA-60079-11:2009.
Non-Incendive for Class I, Division 2, Groups A, B, C, and D.
Dust Ignition-proof for Class II, Division 1, Groups E, F, and G.
Ambient Temperature Limits T4 (T
amb
= -50 °C to 70 °C)
Non-incendive when installed in accordance with Rosemount drawing 00849-1000.
For use with Rosemount power module P/N 753-9220-0001 only.
Enclosure Type 4X / IP66