30
Rosemount 848T Family
November 2014
www.rosemount.com
Analog sensors 4–20mA
Two types of alarm levels are available with 4-20 mA sensors on
the 848T. These types must be ordered with the S002 option
code complete with an analog connector kit. The alarm levels,
accuracy for each type are listed in Table 15.
Table 15. Analog Sensors
Sensor option
Alarm levels Accuracy
4–20 mA
(Rosemount Standard)
3.9 to 20.8 mA ± 0.01 mA
4–20 mA (NAMUR) 3.8 to 20.5 mA ± 0.01 mA
Vue de la page 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 36